ThermoBLOCK
Thermal Coagulation RF Ablation Device

ThermoBLOCK is a system that uses thermal energy to destroy both the endothelium and the collagen within the walls of veins. The ThermoBLOCK catheter heats the vein wall to 120°C. This temperature is sufficient to cause proteins in the wall of the vein to denature.


Share

Catheter Model    Thermoblock Plus
Introducer Sheath
(Minimum ID size)
   6F,7F (2.0 mm)
Insertable Lenght    60, 90, 100cm
Heating Element Diameter    2.0 mm
Heating Element Length    5 cm
Maximum Power Setting    40W
Default Target Temperature Setting    120-130 °C

Horizon RF Ablation Systems probe is a sterile, single-use disposable, packed in separate laminated sterile packaging. The system consists of a generator device that generates radio waves and needles through which it is transmitted to the tissue. In principle, radio waves transmitted through needles inserted into the skin cause the capillaries to coagulate by heat. Small veins on the face or legs are treated with a needle holder combined with specially designed thermo-coagulation needles

Detailed specifications

ThermoBLOCKEndovenous RF Ablation intended for endovascular coagulation of blood vessels in patients with superficial vein reflux. The catheter is provided sterile and is a single-use, disposable device.

ThermoBLOCK is a system that uses thermal energy to destroy both the endothelium and the collagen within the walls of veins.ThermoBLOCK catheter heats the vein wall to 120°C. This temperature is sufficient to cause proteins in the wall of the vein to denature. Thermal injury results in changes in the collagen of the vein wall and contraction of the vessel such that no blood can flow through it. Over time, the vein is gradually reabsorbed by the body’s natural mechanisms so that there is little if any risk of recanalization.

SCIENTIFIC EVIDENCE

 

  • 6F 100 CM
  • 5 cm active tip
  • Rapid and uniform closure of the vein
  • Laser guide guidance
  • Temperature 80-130 degrees 
  • Operation time 20 seconds
  • Easy setup, RF generator
  • Consistent & effective temperature-controlled energy delivery
  • Adjustable temperature and time interval selections are available
  • Intravascular heat parameters are monitored in real-time
  • High catheter echogenicity and better visibility under USG
  • Easy repositioning between segments
  • Fully disposable Single-use RF Ablation System
  • Segmental procedure
  • Tumescent anesthesia
  • Less pain and bruising than laser energy treatment
  • Post-procedure compression stockings required

 

Catheter Model    Thermoblock Plus
Introducer Sheath
(Minimum ID size)
   6F,7F (2.0 mm)
Insertable Lenght    60, 90, 100cm
Heating Element Diameter    2.0 mm
Heating Element Length    5 cm
Maximum Power Setting    40W
Default Target Temperature Setting    120-130 °C
Order Number Device outer diameter Working Length Heating Element Length
VS1001 6.0 Fr 60 cm 5 cm
VS1002 6.0 Fr 90 cm 5 cm
VS1003 6.0 Fr 100 cm 5 cm
VS1004 7.0 Fr 60 cm 5 cm
VS1005 7.0 Fr 90 cm 5 cm
VS1006 7.0 Fr 100 cm 5 cm